The separation is performed in a column (containing either a solid or liquid stationary phase) that has a continuous flow of mobile phase passing through it [usually an inert carrier gas, but more recently supercritical fluids (SCFs) have been used for some applications], maintained in a temperature–regulated oven. When a mixture of substances is injected at the inlet, each component partitions between the stationary phase and the gas phase as it is swept towards the detector. Molecules that have greater affinity for the stationary phase spend more time in that phase and consequently take longer to reach the detector. The detector produces a signal proportional to the amount of substance that passes through it, and this signal is processed and fed to an integrator or some other recording device. Each substance that elutes from the column has a characteristic retention time, defined as the time interval from injection to peak detector response. Fig.1 shows a schematic of a GC system.
Figure 1. A modern GC system
Identification of components was traditionally based primarily on peak retention time, but it is becoming increasingly more reliant on the nature of the response obtained from the detector. The analyst has two main goals: firstly, to make each different compound appear in a discrete band or peak with no overlap (or co–elution) with other components in the mixture, and, secondly, to make these bands uniform in shape and as narrow as possible. This is achieved partly by judicious choice of the column stationary phase and its loading, and partly by optimising the operating conditions of the column. In addition, the method of introducing the sample into the chromatograph, the detector choice and chemical modification to improve the volatility of the compounds also contribute to converting a mediocre analysis into a first–class one.
There has been a continuous synergism between enhanced detector performance and column performance; each advance being mutually dependent on the other. Initially, high–sensitivity detectors permitted the development of a precise column theory that, in turn, enabled the design of columns that had much higher efficiencies. This improved efficiency, however, produced peaks of small volume compared to the sensing volume and dispersion that occurred in the collecting tubes of the contemporary detectors. Consequently, efficiency became limited by the geometry of the detector, and not by its intrinsic sensitivity. Detector design was modified to incorporate smaller tube dimensions, and the volume of the sensing cell was thus reduced greatly. The introduction of capillary columns suitable for routine use in GC in the mid–1980s provided much higher resolution efficiencies and smaller peak volumes, which provoked further modifications in detector design. The latest column developments mean that peaks of only a few milliseconds are a real possibility in both GC and liquid chromatography (LC). Since this matches the response time of the current sensor electronics, the ingenuity of the detector design engineers is again being tested.
Gas chromatography columns
Solid stationary phases
Graphitised carbon black
Carbopaks are graphitised carbon black, having adsorptive surfaces of up to 100 m2/g. They are usually modified with a light coating of a polar stationary phase. Difficult separations of the C1 to C10 hydrocarbons can be achieved rapidly. Carbopak C with 0.2% Carbowax 20M has been used to resolve substances abused by ‘glue–sniffers’. Carbopak C modified with 0.2% Carbowax 1500 and Carbopak C with 0.8% tetrahydroxyethylene diamine (THEED) are useful for the analysis of ethanol and ethylene glycol, respectively, in blood. Resolution is superior to that obtained with the Porapak and Chromosorb polymers, although the elution order is similar.Molecular sieves
Activated alumina is unique for its extremely wide pore–diameter range, and is very useful for separating most C1 to C4 molecules, light hydrocarbon saturates from unsaturates in the C1 to C5 range, and for separating benzene, toluene and xylenes. Deactivation of alumina with potassium hydroxide reverses the elution of some molecules (acetylene and n–butane).Carbosieves are granular carbon molecular sieves that give good separation of C1 to C3 hydrocarbons. Carboxen 1006 is useful in resolving formaldehyde, water and methanol, and impurities in ethylene. Zeolites (5A, 13X) give a good general separation of inorganic gases. Carbon dioxide is irreversibly adsorbed below 160°. Oxygen, nitrogen, carbon monoxide and methane are well separated. These columns have a tendency to adsorb water and carbon dioxide, which results in changes in retention over time.Polymers
Chromosorb 101–108 and Porapak are divinylbenzene cross–linked polystyrene copolymers. Incorporation of other functional groups, such as acrylonitrile and acrylic esters, into the polymer matrix provides moderately polar to polar surfaces with different pore sizes and surface areas (polarity increases with ascending number or letter). HayeSep phases are polymers of divinylbenzene and ethylene glycol dimethylacrylate. Separations range from free fatty acids to free amines, and small alcohols from methanol to pentanol. Tenax-TA is a porous polymer of 2,6–diphenyl–p–phenylene oxide, used both as a chromatographic phase and as a trap for volatile substances prior to analysis. It is also used for high–boiling alcohols, polyethylene glycols (PEGs), phenols, aldehydes, ketones, ethanolamines and chlorinated aromatics.Liquid stationary phases
In GLC, the stationary phase is a liquid or gum at the normal operating temperature. Components injected into the column are partitioned between the moving (mobile) gas phase and a stationary phase. Molecules that have greater affinity for the stationary phase spend more time immobilised in the column and consequently take longer to reach the detector. The process of immobilisation and subsequent release back into the mobile phase occurs thousands of times during the course of the analysis. The separation of components is dependent, to a large extent, on the chemical nature of the stationary phase. Stationary phases are essentially two types of high–boiling polymers, siloxanes (often incorrectly called silicones) and PEGs. Chiral stationary phases based on cyclodextrins (cyclic glucose chains) have specific applications in the separation of enantiomers and are discussed separately.Polysiloxanes
Standard polysiloxanes (PSXs) are characterised by their repeating siloxane backbone in which each silicone atom has the potential to attach two functional groups, the type and amount of which distinguish the stationary phase and its properties (Fig. 2). The basic PSX is 100% methyl–substituted, so when other groups are present, the amount is indicated as a percentage of the total substituent groups. For example, if 5% of Si atoms contain two phenyl groups and the remaining 95% of Si atoms are methyl–substituted, this may be written as (5% diphenyl/95% dimethyl)-PSX, (5% phenyl/95% methyl)-PSX or simply (5% phenyl)-PSX. In some instances, two different groups are present on the same silicone atom, so a (10% cyanopropylphenyl/dimethyl)-PSX contains a total of 5% cyanopropyl, 5% phenyl and 90% methyl residues (see Fig.2). While PSXs are generally less polar in nature than PEGs, the substitution of polar residues for a proportion of the methyl groups confers added polarity to the column. Polar phases retain polar compounds more effectively than do non–polar compounds, and vice versa. The 100% methyl–substituted-PSX is often considered the ‘standard’ non–polar phase, and has been used extensively in compilations of retention indices. This column is an ideal choice for starting a new application. However, substitution by n–octyl groups (up to 50%) renders the column extremely non–polar, and similar to squalene. Substitution with up to 5% phenyl groups still furnishes an essentially non–polar column with improved thermal stability. This phase has also been used for retention index (RI) work and is another good column with which to start a new application. Increasing the phenyl substitution to 20, 35 or 50% yields columns classed as intermediate in polarity, which predictably retain aromatic compounds relative to aliphatic solutes. All these are available as bonded phases that can be solvent rinsed, are not damaged by organic acids or bases, and can tolerate small injections of water if sufficiently highly loaded, but are sensitive to strong inorganic acids and bases. Substitution of cyanopropylphenyl groups (typically 6% or 14%) creates an intermediate polarity mixed phase with unique elution relative to simple phenyl substitution, but renders the column more susceptible to damage from oxygen, moisture and mineral acids. 50% substitution is specifically designed to separate cis- and trans- fatty acid methyl esters (FAMEs). However, even low–level bleeding of the stationary phase produces a high background signal with certain types of detector (nitrogen–phosphorus detector). More polar columns are produced by substitution of bis–cyanopropyl and cyanopropylphenyl groups (80:20 or 90:10). 100% biscyanopropyl substitution gives the highest polarity of the PSXs. This phase can be operated at both high and low temperatures. To date, these are non–bonded columns, and should not therefore be rinsed with solvent. Trifluoropropyl/methyl-PSX is a mid–to–high polarity phase especially suited for otherwise difficult–to–separate positional isomers. Its unique interactions with nitro, halogen, carbonyl and other electronegative groups give it application in the analysis of herbicides and pesticides. Increasing polarity in general is associated with some negative effects. Polar phases tend to have a narrower operating temperature range (higher minimum, lower maximum), are more prone to bleed at higher temperatures, more sensitive to moisture and oxygen, and consequently have a shorter life expectancy than non–polar phases. More recently, low–bleed arylene stationary phases (sometimes designated mass spectrometry or ‘ms’) have been introduced that have phenyl groups incorporated into the siloxane backbone (see Fig. 2). The incorporated phenyl groups confer additional strength to the backbone, which prevents the formation of cyclic fragments and associated ‘bleed’ at higher temperatures.Polyethylene glycols
PEGs are widely used polar stationary phases and their general structure is in Fig. 2. They are less stable, less robust (especially to oxidative damage), have lower temperature limits and a shorter life expectancy than polar PSX phases, but they have unique resolving qualities. Acid–modified PEG (FFAP) substituted with terephthalic acid is especially useful to separate acidic polar compounds, such as acids, alcohols, acrylates, ketones and nitriles. Nitroterephthalic acid–substituted PEG (e.g. Nukol) is designed for volatile fatty acids and phenols. Both are highly resistant to damage from water–based samples. Base–modified PEG (CAM) is suited to analysis of strongly basic compounds, such as primary amines, that do not chromatograph well on polar PSXs. Since this phase is usually cross–linked, it cannot be used with water or alcohol, but it can be solvent rinsed. New bonded and cross–linked PEG phases are now available to separate free fatty acids and other organic acids. These show superior inertness and can tolerate repeated injections of water, alcohols, aldehydes and acids without the need for acidification treatment.Chiral phases
Second–generation chiral phases are based on cyclodextrins (toroidal shaped structures formed by α1–4 linkages of multiple glucose units). The enzyme cyclodextrin glucosyl transferase is used to cleave partially digested starch, and link the glucose units into three forms, referred to as α, β, and γ, that have six, seven and eight glucose units, respectively. The mouth of the cyclodextrin molecule has a larger circumference than the base and is linked to secondary hydroxyl groups of the C2 and C3 atoms of each glucose unit. The primary hydroxyl groups are located at the base of the torus, on the C6 atoms. The number of glucose units thus determines the cavity size and electrophilic orientation (see Fig. 2), and affects the order of the enantiomeric forms. The hydroxyl groups can be functionalised selectively to provide various physical properties and inclusion selectivities. Six different cyclodextrin derivatives are manufactured: permethylated hydroxypropyl (PH), dialkylated (DA), trifluoroacetylated (TA), propionylated (PN), butyrylated (BP) and permethylated (PM). Changes in elution order can be seen between the different derivatives, and also between cyclodextrin cavity sizes. Unlike the cyclodextrins used in high performance liquid chromatography (HPLC), these phases separate both aromatic and non–aromatic enantiomers of a wide range of chemicals, including saturated alcohols, amines, carboxylic acids, epoxides, diols, lactones, amino alcohols, amino acids, esters, pyrans and furans. Each phase has an area of specificity, as summarised in Table .1. Derivatised cyclodextrins are thermally stable, highly crystalline and virtually insoluble in most organic solvents. Chiral phases are fragile, however, and unless chemically bonded or cross–linked, they cannot be washed with solvent or taken to temperatures outside the 0 to 225° range. A subsequent development has been the embedding of PM cyclodextrins (usually 10 or 20% by volume) into columns that contain standard liquid stationary phases of intermediate polarity, such as 35% phenyl-PSX. Silyl–substituted cyclodextrins, such as 2,3–di-O-methyl–6-O-t–butyldimethylsilyl are also available embedded (usually 25 to 35% by weight) in 20% phenyl-PSX, another intermediate polarity stationary phase. These columns are useful to separate positional isomers (phenols, xylenes, etc.), as well as enantiomers.Solute–stationary phase interactions
For liquid stationary phases, three major types of interaction determine chromatographic elution: dispersion, dipole and hydrogen bonding. Table.2 shows the contribution of each of these interactions for the common types of liquid stationary phases. Dispersion is the dominant interaction for all PSX and PEG stationary phases; it can be simplified into the concept that the more volatile the compound (the lower its boiling point), the more likely it is to be in the mobile phase and so the faster it elutes from the column. Although this holds true for groups of compounds with similar functional groups or within homologous series, it cannot be applied universally. In general, a difference of 30° in boiling point is sufficient to predict and maintain elution order, but differences of less than 10° can be overturned by the influence of other interactions. Dipole interactions of PEG phases and the cyanopropyl- and trifluoropropyl–substituted PSXs enable these phases to separate solute molecules that have different dipole moments. Such solutes are those with positional isomers of electronegative groups, such as pesticides, halocarbons and many pharmaceuticals. Moderate hydrogen bonding is exhibited by PEGs and cyanopropyl–substituted PSXs, with less marked effects shown by phenyl- and trifluropropyl–substituted PSXs. Functional groups that exhibit strong hydrogen bonding include alcohols, carboxylic acids and amines; aldehydes, esters and ketones generally have less effect; hydrocarbons, halocarbons and ethers produce negligible hydrogen bonding. Although the amount of separation obtained through dipole interactions or through hydrogen bonding can be difficult to predict, resolution of compounds with smaller differences in dipoles or in hydrogen bonding strengths requires larger percentages of siloxane substitution.McReynolds constants
The retention behaviour of five probe compounds (benzene, butanol, pentan–2–one, nitropropane and pyridine) has been used traditionally to classify stationary phases in terms of their polarity (McReynolds 1970). The retention indices of each of these five reference compounds are measured on the stationary phase being tested, and then compared to those obtained under the same conditions on squalene (a standard non–polar phase). The differences in the retention indices between the two phases (ΔI) for the five probe compounds are added together to give a constant, known as the McReynolds constant, which is used to compare the ability of stationary phases to separate different classes of compounds. Phases that provide McReynolds values of ±4 can be substituted freely for each other; those differing by ±10 units generally yield similar separations. Phases with McReynolds values below 100 are considered non–polar, those above 400 indicate a highly polar phase and values between 100 and 400 an intermediate polarity. Table .3 shows the McReynolds constants, operating temperature range, the relationship between capillary and packed column nomenclature and example applications for the most popular stationary phases. ΔIs for individual probes indicate the deviation from boiling point order and consequently represent the contribution of forces other than dispersion to elution for that probe. The probes are chosen to represent different functional groups as follows:- Benzene for aromatics and olefins (pi–type interactions).
- Butan–1–ol for alcohols, nitriles, carboxylic acids and diols (electron–attracting effect).
- Pentan–2–one for ketones, ethers, aldehydes, esters, epoxides and dimethylamino derivatives (dipole–dipole effect).
- Nitropropane for nitro and nitrile derivatives (electron–donating effect).
- Pyridine for bases (non–bonding electron attraction and hydrogen–bonding effects).
Capillary phasea | Packed equivalent | Temp range (min/max) | McReynolds valuesb | ΣΔI | Applications | ||||
---|---|---|---|---|---|---|---|---|---|
x′ | y′ | z′ | u′ | s′ | |||||
SPB-Octyl | Squalene, Apiezon L | –60/300 | 3 | 14 | 11 | 12 | 11 | 51 | Separates by boiling point, polychlorinated biphenyls (PCBs) |
*-1 | SE30, OV-1, OV-101, SP2100 | –60/320 | 4 | 58 | 43 | 56 | 38 | 199 | Amines, hydrocarbons, pesticides, PCBs, phenols, sulfur compounds, flavours, fragrances |
*-5 | SE-52, OV-73 | –60/320 | 19 | 74 | 64 | 93 | 62 | 312 | Alkaloids, drugs, FAMEs, halogenated compounds, aromatic compounds |
*-1301 | SE-54, CP-624 | –20/280 | 69 | 113 | 111 | 171 | 128 | 592 | Aroclors, alcohols, phenols, volatile organic acids |
*-35 | OV-11 | 0/300 | 101 | 146 | 151 | 219 | 202 | 728 | Aroclors, amines, pesticides, drugs |
*-1701 | OV-1701 | 10/280 | 67 | 170 | 153 | 228 | 171 | 789 | Aroclors, herbicides, pesticides, trimethylsilyl (TMS) sugars |
*-50, *-17 | OV-17, SP-2250 | 30/310 | 125 | 175 | 183 | 268 | 220 | 971 | Drugs, glycols, pesticides, steroids |
*-210 | OV-25 | –45/250 | 178 | 204 | 208 | 305 | 280 | 1175 | Aldehydes, ketones, organochlorines, organophosphates |
*-225 | XE-60, OV-210 | 40/230 | 146 | 238 | 358 | 468 | 310 | 1520 | FAMEs, alditol acetates, neutral sterols |
*-23 | OV-225 | 40/250 | 228 | 369 | 338 | 492 | 386 | 1813 | cis–transFAMEs, stereoisomers |
*-wax, *-20M | Carbowax 20M TPA | 35/280 | 305 | 551 | 360 | 562 | 484 | 2262 | Alcohols, free acids, essential oils, ethers, glycols, solvents, primary amines |
*-FFAP | Carbowax–1500 PEG | 50/250 | 340 | 580 | 397 | 602 | 627 | 2546 | Acids, alcohols, aldehydes, acrylates, nitriles |
Nukol | SP-1000, OV-351 | 60/200 | 314 | 569 | 372 | 578 | 504 | 2337 | Alcohols, free acids, essential oils, ethers, glycols, solvents |
*-2330 | SP-2330 | 10/250 | 382 | 610 | 506 | 710 | 591 | 2799 | cis–transFAMEs, positional isomers |
*-2380 | – | 10/275 | 402 | 629 | 520 | 744 | 623 | 2918 | cis–transFAMEs, positional isomers, alditol acetates |
*-2340 | SP-2340, OV-275 | 25/250 | 419 | 654 | 541 | 758 | 637 | 3009 | cis–transFAMEs, positional isomers |
– | DEGS | 20/200 | 496 | 746 | 590 | 837 | 835 | 3504 | Acids, esters, phenols, terpenoids |
– | EGS | 100/200 | 537 | 787 | 643 | 903 | 889 | 3759 | TMS or methyl sugars, acidic drugs |
TCEP | TCEP | 10/145 | 594 | 857 | 759 | 1031 | 917 | 4158 | Flavours, fragrances, essential oils |
α-Cyclodextrin in 35% phenyl-PSX | 30/240 | 102 | 243 | 142 | 221 | 170 | 878 | Enantiomers and isomers (see Table .1) | |
β-Cyclodextrin in 35% phenyl-PSX | 30/240 | 119 | 264 | 154 | 134 | 187 | 858 | Enantiomers and isomers (see Table .1) |
Solid supports for packed columns
The raw material for the most commonly used supports is diatomaceous earth, calcined, usually with a flux, then crushed and graded into a number of particle sizes (60 to 80, 80 to 100 and 100 to 120 mesh). Chromosorb G has a surface area of about 0.5 m2/g, and is suitable for low–loaded columns – the amount of stationary phase should not exceed 5% (w/w). Chromosorb W has a larger surface area (1 m2/g) and accepts higher loadings, but is more fragile. Chromosorb P, obtained from crushed firebrick, accepts loadings up to 35% (w/w) for some phases. Supelcoport is the most inert of the diatomite supports, and can accept 20% (w/w) loadings. Carbopack, Porapak, HayeSep and Tenax-TA (see above) can also be used as solid supports. Table.4 shows the maximum suggested coating percentages for the most common solid supports. Deactivated support materials are nearly always preferred. Deactivation procedures include acid or base washing to remove impurities and fines, and treatment with a silanising agent that reacts with surface hydroxyl groups to reduce adsorptive effects. Support materials that have the liquid phase chemically bonded to them are available. These offer decreased bleed rates of stationary phase, an advantage when operating a temperature programme or when using a mass spectrometer as the detector.Installing, conditioning and maintaining columns
Column installation
Glass columns must have straight smooth ends to allow them to fit into unions at the injector and detector using either graphite or Vespel (polyimide) ferrules inside the nuts. Fused silica capillaries should also have their ends freshly cut after insertion through ferrules, to eliminate blockages. The injector end of the column should be fitted first, adjusting the height of the protrusion above the ferrule according to the type of inlet being used, then tightening the fittings just enough to prevent leakage when tested with a proprietary leak–testing fluid (not soap solution. which leaves a residue). The detector end of a packed column can be attached to a bubble–flow meter to ensure adequate flow through the column before connecting the detector end. The detector end of the capillary column may be immersed into a small tube of methanol to ensure adequate flow, the capillary end re–cut, then attached to the detector and checked for leaks. The detector is activated and the column tested at room temperature with an injection of 1 or 2 μL of methane, when a needle–sharp peak should be obtained. When the carrier-gas pressure has been adjusted to give a flow of approximately 20 mL/min of carrier gas for a packed column or 1 to 2 mL/min for a capillary, the column may be heated and a test mixture injected. Commercial columns are invariably supplied with a chromatogram obtained from a test mixture, and it should be possible to obtain a performance equal to or better than the supplied chromatogram. Various test mixtures are used, including a mixture of dimethylphenol and dimethylaniline with straight–chain paraffins. Any acidity or alkalinity of the column is apparent by loss of the peak shape of the amine or phenol. The efficiency obtained is a function of the entire chromatographic system. Poor efficiency or peak shape often results from a non–swept volume somewhere in the system. It may be necessary to add an additional gas supply to the column outlet to ensure that the detector is purged effectively, because most detectors are designed to operate with packed columns and a flow rate of about 30 mL/min, as opposed to the 1 or 2 mL/min delivered by a capillary column (see Dectectors below).Column conditioning
A new column requires conditioning before use to remove volatile impurities that remain after deactivation of the support, and/or the coating and packing processes. The column should be installed in the injector port only, with the detector end disconnected. With the column at room temperature, a low carrier gas pressure (14 to 35 kPa) should be maintained for half an hour to purge oxygen from the system. The temperature may then be raised by about 1°/min, until about 10° above the desired maximum operating temperature has been reached, and the column is maintained (2 h for a manufactured capillary, 12 h for a packed column or self–packed capillary). Care must be taken not to exceed the maximum operating temperature. After conditioning, the column is connected to the detector, and a period of further conditioning instigated if the background signal is excessive. Some phases (e.g. OV-17) are particularly oxygen–sensitive and can be ruined by careless conditioning. A constrictor fitted to the detector end of the column helps prevent back diffusion of oxygen if air or oxygen is supplied to the detector.Guard columns and retention gaps
A guard column and a retention gap are essentially the same thing, but are installed to serve different purposes. These 1 to 10 m lengths of fused silica tubing are attached to the front of the chromatography column via a press–snap connector or zero dead–volume union, and then installed into the injector port. The surface of the silica is deactivated to minimise solute interactions, but no stationary phase is added. The diameter tubing should be the same as the column, but if different it should ideally be of a wider bore. The function of a guard column is to trap deposits of non–volatile residues preventing their contamination of the analytical column. Solutes are not retained by the guard column (since there is no stationary phase) and pass directly onto the column. Portions can be cut periodically from the top of the guard column as deterioration in chromatography requires, without any appreciable loss of resolution from the analytical column. A retention gap is used to improve peak shape when poor chromatography is the result of either a large injection volume (>2 μL), or there are solvent–stationary phase polarity mismatches. Greatest improvement is seen in early eluting peaks, or for solutes with similar polarity to that of the solvent.Maximum operating temperatures
Maximum operating temperatures for stationary phases are usually quoted assuming isothermal operation with a flame ionisation detector (see Table .3). Other detectors may impose different limits, the mass spectrometer being much more susceptible to bleeding of the stationary phase than the thermal conductivity detector. All phases bleed very slightly at high temperatures because of the loss of smaller sized (and hence lower boiling) polymer chains, although normally this is not noticeable. Operating temperature has a profound effect on column life, particularly for capillary columns. Loss of stationary phase, or breakdown of the thin film into pools to expose part of the tubing surface, results in serious loss of performance. Additionally, in columns that contain PSX phases with two different functional groups, one group (usually that which confers additional polarity) is preferentially lost. This results in a change of relative separation (or in RI– see discussion below) as well as a loss of resolution. The temperature limit of a column may be determined by the deactivation procedure used in production, rather than by the stationary phase itself. Newer silica columns have a very low metal oxide content, thought to act as a catalyst for the degradation of both sample and stationary phase, and thus enable phases to be run at higher temperatures. Fused silica capillary columns have a protective external coating of polyimide that is slowly degraded at elevated temperatures (maximum temperature 360°, but now to 400°), which can also limit column life. However, separations are usually achieved at much lower temperatures.Temperature programming
For complex mixtures with components of widely varying retention characteristics, it is often impractical to choose a column temperature that allows all the components to be resolved. Increasing the column temperature throughout the analysis dramatically reduces the time taken for higher boiling compounds to elute, and simultaneously improves the sensitivity of the assay, as the peaks are remarkably sharper. If the early eluting compounds are resolved inadequately, a lower starting temperature or slower initial ramp should be used, taking care to observe the temperature requirements of the type of injector used. All instruments currently manufactured are available with a temperature programme option, and a multi–ramp programmer is particularly useful for capillary chromatography. The first ramp can be used during splitless injection (see later) to bring the column rapidly up to the initial chromatography temperature, followed by a slower analytical ramp to perform the separation. One problem with temperature programming is that the backpressure increases with temperature and reduces the carrier gas flow if a mass–flow controller is not used. For polar stationary phases, the polarity increases with temperature, which causes distortion of RI data (see discussion on the use of RIs below). Column bleed also increases, which results in an increasing baseline. For this reason the column should be well conditioned before use.Evaluating column performance
Retention time
Retention time (tR) is the time taken for a given solute to travel through the column, and is the time assigned to the corresponding peak on the chromatograph. It is a measure of the amount of time the solute spends in the column, and is therefore the sum of time spent in both the stationary and the mobile phases.Retention time of a non–retained compound or hold–up time (tM or t0)
The retention time tM is the time taken for a non–retained solute to travel along the column; it represents the transit time for the mobile phase (carrier gas) in the column and is a column–specific parameter, applicable only under the prevailing conditions of gas flow and oven temperature. It is the same for all solutes on the column, and no other peak can be expected to elute earlier than this time. tM is obtained by injecting a non–retained compound suitable for the detector system being used [butane or methane for flame ionisation detection (FID) or thermal conductivity detection (TCD), acetonitrile for nitrogen–phosphorus detection (NPD), methylene chloride for electron–capture detection (ECD), vinylchloride for photoionisation detection (PID) or electrolytic conductivity detection (ELCD)].Average linear velocity
The average linear velocity (μ̄) represents the average speed of carrier gas through the column, usually expressed in cm/s, and is considered more meaningful than measuring the flow (usually expressed in mL/min) at the column effluent, since flow is dependent on column diameter. This term directly influences solute retention times and column efficiency. Velocity is controlled by altering the column head pressure, and is calculated from the equation
Retention factor
Retention factor (k) is the ratio of the amount of time a solute spends in the stationary and mobile phases and is calculated from tR and tM using the equation:Separation factor
The separation factor (α) is a measure of the time interval between two peaks. If α equals 1, then the peaks have the same retention time and co–elute. Separation factor is calculated using the equation:Number of theoretical plates or column efficiency
The theoretical plate is an indirect measure of peak width at a specific retention time. Higher plate numbers indicate greater column efficiency and narrower peaks. The number of plates per metre of column (N) is calculated from either form of Equation (28.4): N = 16(tR/wb)2(28.4a) N = 5.54(tR/wh)2(28.4b) where tR is the time from injection to peak maximum for the solute, wb is the peak width at base in units of time and wh is the peak width at half height in units of time. Efficiency is a function of the column dimensions (diameter, length, film thickness or loading), the type of carrier gas and its flow, as well as the chemical nature of the solute and the stationary phase.Separation (Trennzahl) number
Trennzahl formulated this measure of how many resolved peaks can be accommodated between adjacent members of a homologous series. The separation number for A and B, two members of a homologous series that differ by one methylene group, is given by:Peak shape or asymmetry
A well–designed system should give symmetrical peaks, as tailing or fronting adversely affects resolution. Tailing may result from non–swept volume in the system or from component–stationary phase or component–support interactions. Tailing of polar compounds can often be remedied by the use of a more polar stationary phase. Fronting (shark’s fin peaks) is usually caused by overloading, particularly with capillary columns, and can be resolved either by making a smaller injection or by using a column with a higher stationary phase ratio. Column capacity is the maximum amount of a solute that can be chromatographed successfully without loss of peak shape. Table .5 shows the relationship between column capacity, film thickness and column diameter. Peak fronting caused by thermal decomposition can be reduced by either lowering the injection temperature or using a cold on–column injector system. Peak shape is usually expressed by the peak asymmetry (As). In Fig. 3, the peak asymmetry factor for substance B is given by:
Use of retention indices
Inlet systems
Megabore direct (or packed column) injection
Split and splitless injectors
Cold on–column injection
Cold on–column injection is most suited to compounds that are thermally labile. The injection needle must be fine enough to enter the column bore, usually fused silica or stainless steel with a fused silica insert. The top of the column is held at a temperature low enough for the solvent that contains the sample to condense, usually by an air- or carrier gas–cooled sleeve. The solvent temporarily swamps the stationary phase and ensures that the sample components concentrate in a narrow band. Any solvent or sample that remains in the injector is backflushed with carrier gas, often by automatic valves. The proximal end of the column is then brought rapidly to the operating temperature, when the solvent vaporises and chromatography begins. The potential for rapid column contamination or deterioration means that cold on–column injection is usually restricted to those applications where its use is essential.Volatiles interface
The volatiles interface allows automated analysis of gaseous samples. The interface is a low–volume highly inert switching block, and is ideally suited to trace–level detection. A portion of the carrier gas supply is diverted through the specimen sampler and released under controlled conditions onto the column. The remainder of the carrier gas goes to a flow sensor, which prevents fluctuations in column gas flow that would otherwise occur when the switching valves are opened and closed. The interface can be run in split, splitless or cold–on column modes as described in the sections above. Samples may be introduced from external devices, such as air samplers or purge and trap devices (see section below), or from headspace analysis, which permits analysis of volatile substances in a liquid sample while minimising contamination of the column. This technique is used in the assay of ethanol and other solvents in blood and for complex household preparations, such as polishes, which contain volatile substances.Thermal desorption and purge–and–trap injection
The analysis of samples that have been pre–concentrated onto solid adsorbents is common in the fields of industrial air monitoring, analysis of residues in food, soil and water, petrochemical analysis, and environmental monitoring. The methods of preparing samples for analysis are described in the section on specimen preparation. These samples require special interfaces with GCs to ensure good chromatography. In some instances the sample preparation device and injector are manufactured as stand–alone pieces of equipment that require very little modification of conventional injectors, while others must be dedicated pieces of equipment. Once collected, the concentrated sample must be desorbed into the chromatograph using the heated injector port. The major problem here is the possible introduction of water into the chromatograph from moisture adsorbed during collection from high–humidity samples. Release of solutes from the adsorbent should be as rapid and complete as possible to allow for rapid and sensitive analysis and for a narrow sample band to be introduced into the chromatograph. This is achieved either by cooling the column oven cryogenically to refocus the sample in the injector prior to injection or by using a dry purge system coupled to the GC via a volatile interface (see above) designed to operate above ambient temperature. Here, the specimen is thermally desorbed from the collection tube onto a narrower (1 mm i.d.) tube of the same adsorbent material. The concentrated solute is then released into the chromatograph, ensuring rapid and complete sample introduction. Adsorbents must be thermally stable to reduce interference from background contaminants. With solid–phase micro–extraction (SPME) the adsorbed sample is introduced into the heated injector port via a special sleeved needle (see under specimen preparation). This technique requires the injector liner to be narrow (usually 0.75 mm as opposed to 2 or 4 mm) to increase the linear velocity of carrier gas through the liner and ensure a narrow band of sample is introduced onto the column.Solid injection
When solvent interference is serious the sample may be injected as a solid. The ‘moving needle’ injector has found application in steroid analysis and for the determination of anticonvulsant drugs. A solution of the material to be injected is placed on the tip of the glass needle with a syringe. A small flow of carrier gas sweeps the solvent out of the top of the device to waste. The dry residue is then introduced by moving the needle into the heated injection zone of the chromatograph with a magnet. This form of injection can only be used with compounds that do not volatilise with the solvent.Backflush
Upon vaporisation, the injected sample undergoes considerable expansion, sometimes up to 100 to 1000 times its original volume, which creates a pulse of pressure that often exceeds the column carrier-gas pressure. If the volume of the liner is smaller than the expanded solvent volume (see Table.6), some of the sample is propelled out of the injector in a process known as backflush. This can appear as a broad tailing solvent front, since it now takes longer to flush the expanded solvent out of the injector and carrier-gas line. Backflush can also cause injector contamination, since the analytes condense in the cooler carrier-gas line, from where they may bleed continuously into the injector and cause high background or spurious peaks. Carryover or peak ghosting can occur when the next injection backflushes and carries previously condensed compounds back into the vapour phase and onto the column. Backflush can usually be solved by using a smaller injection volume, a less expansive solvent, a lower injector temperature, a liner with an upper restrictor or a faster carrier gas flow. The use of an adjustable septum purge gas (0.5 to 1 mL/min usually) also decreases the potential for backflush, as components that would normally condense on the cooler septum and travel into the carrier-gas lines are swept away by the septum purge. Too high a purge flow results in loss of highly volatile components.Solvent | Boiling point (°) | Expansion volume (μL) per μL of solventa | Suggested GC oven starting temperature (°) |
---|---|---|---|
Methylene chloride | 40 | 330 | 15–30 |
Carbon disulfide | 46 | 355 | 15–30 |
Acetone | 56 | 290 | 30–45 |
Methanol | 65 | 525 | 40–55 |
n-Hexane | 69 | 165 | 40–60 |
Ethyl acetate | 77 | 215 | 45–65 |
Acetonitrile | 85 | 405 | 55–75 |
iso-Octane | 99 | 130 | 70–90 |
Water | 100 | 1180 | 70–90 |
Toluene | 111 | 200 | 80–100 |
Injector discrimination
Injector discrimination occurs because not all the compounds in the sample vaporise at the same rate. Since the sample remains in the liner for a limited time, this usually results in some loss of higher–boiling solutes. This can be alleviated by increasing the residence time of the sample within the injector, or by using a higher injector temperature or smaller injection volume. However, there is usually a compensatory loss in lower–boiling compounds. Discriminating behaviour can usually be managed by making reproducible injections.Gas pressure and flow control
Detector systems
Flame ionisation detector
Nitrogen–phosphorus detector or alkali flame ionisation detector
Electron–capture detector
The early form of this detector consists of a small chamber with a pair of electrodes and a radioactive source, usually 63Ni, placed close to the cathode to ionise the carrier gas. Potential applied to the electrodes produces a steady background current. Electron–capturing solutes arriving in the chamber remove some of the electrons and reduce the detector current. The response of the detector is therefore a loss of signal rather than an increase, as is given by most other detectors. Although the ECD can be polarised from a suitable low–voltage direct–current supply, it is more sensitive when a pulsed power supply is used, and in modern detectors the polarising pulses are modulated to maintain a constant current. A voltage dependent on the modulation frequency is generated as the output signal. Additional carrier gas is necessary, even with packed columns, to obtain a flow of at least 60 mL/min to purge the detector adequately and avoid peak broadening and distortion. Sensitivity can also be improved dramatically by raising the operating temperature of the detector, and decreasing the makeup gas flow.Fourier–transform infra-red detector
Atomic emission detector
With the atomic emission detector (AED), carrier gas that elutes from the column delivers solutes into a high–temperature helium plasma, where heat energy is absorbed by the constituent elements. In returning to their ground state, energy is emitted as light, the wavelength of which is characteristic for each element. Emitted light is focussed by a quartz lens and spherical mirror onto a diffraction grating, and the dispersed light focussed onto a diode array that is continuously scanned (wavelength usually 170 to 800 nm). Typically, some 15 elements can be monitored simultaneously, and each is plotted against time. The composite chromatogram allows the percentage elemental composition of each peak to be determined. Sensitivity is very good, but the detector is complex and expensive to operate and is not widely used.Mass spectrometer
Ion–trap mass spectrometer
As with other forms of mass spectrometers, EI or chemical ionisation (CI) is used to produce an ion source, but this is focussed into the ion–trap mass spectrometer in pulses rather than continuously. The fundamental difference is that all the solute ions generated over the entire pulse period are trapped in the detector and are then sequentially ejected in increasing mass number from the trap into the electron multiplier. The addition of helium into the trap (133 mPa) contracts the ion trajectory to the centre of the trap, where it is further focussed by the ring electrode, to form dense ion packets that are expelled more efficiently than diffuse clouds, and thus greatly improve resolution. The spectral patterns can be quite different to those produced by mass filter spectrometers, and are often characteristic of the conditions under which the instrument is run, which makes comparison difficult between instruments. However, because the ion collection period is longer, the sensitivity of the ion trap in full scan mode is similar to that obtained in SIM on the average MS. Furthermore, an improved mass range (sometimes up to several thousand amu) gives this type of detector many applications, particularly for quantitative trace analysis, and for higher mass components.Dual detector systems
Specimen preparation
Isolation and concentration
Protein precipitation
Liquid–liquid extraction
Liquid–liquid extraction is the most frequently used method to isolate and concentrate solutes for GC. The pH of the specimen is adjusted to ensure the compounds to be extracted are not ionised (basic for bases, acid for acidic compounds). Bearing in mind that some portion of the aqueous acid or base will dissolve in the solvent, the use of strong mineral acids or alkalis is not advised as this adversely affects column performance. Best results are obtained with acidic buffers (phosphate or acetate) and with ammonium hydroxide or basic buffers (borate), using a 5:1 ratio of solvent to specimen. The solvent chosen should be sufficiently polar to partition the compound of interest without co–extracting excessive amounts of polar contaminants. For more water–soluble drugs, such as β-blockers, the addition of 2 to 10% of a polar solvent (e.g. isopropanol or butanol) is helpful, or solid sodium chloride can be added to ‘salt out’ the analyte. If a derivatisation step is to be carried out subsequently, the use of a solvent compatible with the derivatisation eliminates the need for an evaporation step. Use of solvents with a higher density than the sample (e.g. dichloromethane) can lead to difficulty in isolation of the organic phase. Purification of extracts by back extraction (re–extraction of the analytes from the organic solvent at the opposite pH followed by re–extraction into solvent at the original pH) may be helpful for trace analysis. The use of a small volume of solvent for the final extraction serves as a concentration step without the need for separation and evaporation of the organic phase.Solid–liquid or solid-phase extraction
Solid–liquid extraction uses a polypropylene cartridge with a small amount (200 mg to 3 g) of high capacity (1 to 20 mL) silica–based packing at the base of the reservoir. On introduction of the sample matrix, the compounds of interest are withheld by the packing. Impurities are then rinsed selectively from the column, and the final elution releases the compound of interest. Evaporation followed by reconstitution in a suitable solvent provides a clean concentrated sample ready for analysis by GC. Bonded phase packings that have been modified by the addition of various functional groups are available. The mechanisms of interaction for the matrix, analytes and packings are similar to those in LC. Polar stationary phases retain polar analytes (normal phase) and are eluted with organic solvents, while non–polar stationary phases retain non–polar analytes (reverse phase) and are eluted with aqueous solvents. Ion–pair extraction uses a non–polar stationary phase and polar analyte, with a counter ion added to the sample solution, and allows retention of the (now neutral) analyte by a reversed–phase mechanism. In ion–exchange extraction, the adsorbent surface is modified with ionisable functionalities. Analytes with ionic charges opposite to those on the packing are retained. Solvents that contain counter ions of greater strength are used to elute the analytes of interest from the tube.Solid–phase micro–extraction
SPME requires no solvents or complicated apparatus and can concentrate volatile and non–volatile compounds in both liquid and gas samples. The unit consists of a fused silica fibre attached to a stainless–steel plunger coated with a stationary phase (mixed with solid adsorbents as required). The plunger is inserted through a septum into a vial that contains the sample, and the fibre exposed by depressing the plunger either into the liquid or the headspace for 20 to 30 min. The retracted fibre is inserted into the injection port of the GC, and is desorbed when the plunger is depressed. The unit may be reconditioned and used 50 to 100 times. For field analysis, adsorbed samples can be stored and transported in the needle sealed in a special container for subsequent analysis by GC (or LC). Pesticides recovered from water samples have been shown to be more stable when stored in this way than in water. The special small–volume injection liner fits any model of chromatograph, and produces sharper peaks because of the higher linear gas velocity, with little or no backflush. Suitable stationary phases are:- 100 μm dimethyl-PSX film for low molecular weight compounds or volatiles, or a thinner film (7 μm) for larger weight semivolatile compounds
- 85 μm polyacrylate film for polar compounds
- 65 μm film of dimethyl-PSX–divinyl benzene for volatile alcohols and amines
- for surfactants, 50 μm Carbowax–templated resin
- for trace level volatiles, a 75 μm Carbowax–carboxen phase is suitable.
Supercritical fluid extraction
A SCF is a substance that is maintained above its critical temperature and pressure, where it exhibits physicochemical properties intermediate between those of a liquid and a gas. Properties of gas–like diffusivity, gas–like viscosity and liquid–like density combined with a pressure–dependent solvating power provided the impetus to apply SCFs to analytical separation. The initial applications most often involved isolation of flavours and contaminant residues from food and soil. These have now been extended to the isolation of drugs from blood and other aqueous–based media by using adsorbents added in–line (such as molecular sieves, diatomaceous earth, silica gel, etc.) to filter proteinaceous material and adsorb water. It is possible, by adding small volumes of co–solvent to the SCF, to extract highly polar solutes with excellent efficiency. In contrast to the conventional extracting solvent, the fluid most often used in supercritical fluid extraction (SFE), supercritical CO2, is non–polluting, non–toxic and relatively inexpensive. Additionally, extractions are carried out quickly at temperatures that avoid degradation of temperature–sensitive analytes and provide clean extracts with extremely high efficiency. Several dedicated SFE analysers are available; each consists of a gas supply, pump and controller used to pressurise the gas, temperature–controlled oven, extraction vessel, internal diameter regulator and collection device. The CO2 supply is compressed to a selected pressure (e.g. 28 000 kPa) and its temperature adjusted (e.g. 50°). As the supercritical CO2 passes through the sample material, the solutes are extracted to an equilibrium solubility level, typically about 10% (w/w). The gaseous solution that leaves the extractor is passed through the pressure reduction valve, where the pressure (and thus the dissolving power) of the CO2 is reduced. The solutes precipitate in the separator, and the CO2 is recycled through the system several times until the extraction is completed, when it is vented to waste.Headspace analysis
This method of isolation is used for analytes with volatility higher than that of the common extraction solvents.Purge and trap
Purge and trap is a powerful procedure for extracting and concentrating volatile organic compounds from soil, sediment, water, food, beverages, etc. It is especially useful for poorly water–soluble compounds and those with boiling points above 200°. The procedure involves bubbling an inert gas (nitrogen or helium) through an aqueous sample or suspension at ambient temperature, which causes volatile organic compounds to be transferred into the vapour phase. During the purge step, purge gas sweeps the vapour through a trap containing adsorbent materials that retain the volatilised compounds. Water vapour may be removed by dry purging. The trap is rapidly heated to 5 to 10° below the desorb temperature. The valve is then switched to join the trap flow to the carrier-gas flow, and the trap heated to its desorb temperature for a fixed time. Adsorbent tubes are usually packed with multiple beds of sorbent materials, each one more active than the preceding one, which allows compounds with a wide range of boiling points and polarities to be analysed simultaneously. During purge, the smaller and more non–polar solutes are readily carried down the beds, and since the carrier gas passes in the opposite direction during the desorb phase, the larger and more polar compounds do not come into contact with the innermost active beds, from which their release may be difficult to effect.Thermal desorption
This technique is used extensively for air monitoring in industrial hygiene, environmental air, indoor air or source–emission monitoring. The device may be portable or fixed and of varying size. Air is pumped continuously through the device at a fixed rate, during which time components are extracted gradually and concentrated onto the adsorbent beds; the arrangement of the beds is the same as described above for the purge and trap, and prevents potentially irreversible binding of large molecules. The direction of the flow is simply reversed during desorption. Analysis requires a special interface to the GC, which is described above in the section on Thermal desorption. The adsorbents must have high capacity to remain active during the entire sampling period, and show an acceptable pressure drop during sampling. Ideally, a minimal amount of unwanted analytes should be absorbed, as these will contribute to the background noise.Tissues and hair
Tissues and hair require treatment prior to drug extraction to break down the biological matrix and enable a good recovery of the drug. For solid tissues, good results are obtained by incubation of a portion of the tissue with a mixture of a collagenase, a protease and a lipase in a buffer of suitable pH. For small amounts of tissues (100 mg), overnight treatment at room temperature suffices, although gentle agitation or occasional mixing speeds up the process. Larger amounts of tissue benefit from mechanical homogenisation prior to incubation. For the analysis of hair, an initial washing to remove residues from cosmetic products or environmental contaminants is recommended, followed by incubation with either caustic alkali (for basic drugs) or mineral acid (for acidic drugs). After adjustment of the pH, drug recovery can proceed by the usual procedures established for the specific compounds under investigation.Hydrolysis
Recovery of conjugated drug metabolites from biological fluids can be increased by hydrolytic cleavage of the conjugate bond prior to extraction. This offers a vast improvement in sensitivity for qualitative analysis, particularly from urine, and is essential to identify drugs (e.g. laxatives) that are excreted almost exclusively as conjugated metabolites. However, reliable quantitative analysis of conjugated metabolites requires that the unconjugated metabolite must first be removed or quantified, and then the total (conjugated plus unconjugated) metabolite be measured after hydrolysis in a subsequent separate procedure. For quantitative work, appropriate standards that contain conjugated metabolites must be carried through the procedure to monitor the efficiency of the hydrolysis step.Enzymatic hydrolysis
The use of a specific enzyme to cleave chemical bonds is the more specific of the two approaches, but incurs additional cost and time. It also provides cleaner extracts, and therefore prolongs the life of the chromatography column. There are a number of commercial preparations of purified glucurase and sulfatase harvested from different species. It is important to pay attention to the pH and temperature optima of the specific enzyme preparation. Temperature–tolerant preparations allow heating up to 60°, which permits relatively short (2 h) incubation times.Chemical hydrolysis
This quicker and less expensive approach can provide suitable extracts for chromatography for some analytes, although they are generally more demanding in terms of clean–up procedures. Typically, strong mineral acids or alkalis are used, often with boiling or treatment in a microwave or pressure cooker. Extracts must be neutralised, otherwise the chromatography column deteriorates quickly. Care should be taken to ensure the stability of the analytes to the hydrolysis conditions. Vigorous hydrolysis conditions often yield undesirable by–products, or if several compounds can be hydrolysed to a single entity, preclude accurate identification of the original compound present. For example, both the acid and the enzymatic hydrolysis of benzodiazepines remove glucuronide conjugates, but acid hydrolysis also converts two or three drugs to the same benzophenone compound (diazepam, temazepam and ketazolam are all converted into 2–methylamino–5–chlorobenzophenone). While this compound has good chromatography characteristics, the approach is unsuitable for those applications (such as forensic analysis) that require absolute identification of the drug ingested.Derivative formation
Chiral separations
Chiral compounds can be derivatised to improve their chromatographic characteristics, and the enantiomers separated on a chiral stationary phase. Both enantiomers behave similarly, provided that steric hindrance does not preclude a reaction with one enantiomer. An alternative approach is to use a chiral derivatising reagent which, when reacted with enantiomers, produces diastereoisomers that can then be separated on a conventional stationary phase. As with enantiomers, diastereoisomers still produce similar mass spectra, but are resolved in time by the chromatography column. This approach is less expensive and also less restrictive, since a dedicated column is not required. Care should be taken to ensure the enantiomeric purity of the derivatising reagent, and to guard against racemisation during the reaction. n-Trifluoroacetyl–l–propyl chloride (TPC) in triethylamine and chloroform (or ethyl acetate) is a commonly used chiral reagent that couples with enantiomeric amines. Excess reagent is washed off with 6M HCl and the organic phase dried over magnesium sulfate. For chiral alcohols, (1R,2S,5R)-(–)-menthylchloroformate (MCF) reacts well if pyridine is used as a catalyst.Quantitative determinations
Quantitative work usually requires some form of sample preparation to isolate the drug from the bulk of the sample and some degree of concentration or, more rarely, dilution. These processes inevitably introduce a degree of analytical error. A further difficulty is caused by the non–reproducibility of injected volumes. To compensate for these errors, it is usual to compare the response of the unknown with the response of an added internal standard. The internal standard should be added as early as possible in the assay process and should have chromatographic properties matching the drug's as closely as possible, preferably with a longer retention time. It is often possible to obtain unmarketed analogues of drugs, or compounds specially synthesised for use as internal standards (e.g. a methyl addition or a halogen substitution). However, the internal standard usually does not behave exactly as the drug and careful control of variables, such as pH, is necessary. If a derivative is to be prepared, the internal standard should also be amenable to derivatisation. Use of an inappropriate internal standard can seriously affect precision (Dudley 1980). If a mass spectrometer is being used as the detector, then the ideal internal standard is a 3H or 13C substituted analogue of the drug, a number of which are readily available at reasonable cost. Calibration should include points of higher and lower concentrations than the sample, and quality assurance samples should be included at appropriate concentrations in frequently run assays. Peak measurement may be by peak height or by the peak area obtained by integration. If the peaks show even a modest degree of tailing, use of peak area usually provides a more accurate quantitative result. A plot of the ratio of peak height (or area) of the drug to internal standard versus concentration is a straight line with most detectors. Care should be taken in the preparation of standards to match the matrix to that of the specimens, and to allow for any associated salt or water of crystallisation in the calculation of the concentration. The best results are obtained when the amount of internal standard used produces a peak response ratio of one at the mid–point of the calibration range.Optimising operation conditions to customise applications
Additional sensitivity can be achieved by increasing sample size, using a concentration step, derivatisation, injecting a larger sample volume, selecting a different stationary phase and using the detector at a higher sensitivity level.Film thickness d (μm) | Column internal diameter (mm) | ||||||
---|---|---|---|---|---|---|---|
0.10 | 0.18 | 0.20 | 0.25 | 0.32 | 0.45 | 0.53 | |
0.10 | 250a | 450 | 500 | 625 | 800 | 1125 | 1325 |
0.18 | 139 | 250 | 278 | 347 | 444 | 625 | 736 |
0.25 | 100 | 180 | 200 | 313 | 400 | 450 | 663 |
0.40 | 63 | 113 | 125 | 156 | 200 | 282 | 331 |
0.42 | 107 | 119 | 149 | 190 | 265 | 315 | |
0.50 | 90 | 100 | 125 | 160 | 225 | 265 | |
0.83 | 60 | 75 | 96 | 136 | 160 | ||
0.85 | 59 | 74 | 94 | 133 | 156 | ||
1.00 | 50 | 63 | 80 | 113 | 133 | ||
1.27 | 49 | 63 | 88 | 104 | |||
1.50 | 42 | 53 | 75 | 88 | |||
2.55 | 25 | 31 | 44 | 52 | |||
3.00 | 21 | 27 | 38 | 44 | |||
5.00 | 13 | 16 | 23 | 27 | |||
Efficiency Nb | 12 500 | 6600 | 5940 | 4750 | 3710 | 2640 | 2240 |
Specific applications
The systems given below are applicable to the routine screening, separation and identification of groups of drugs and chemicals. They are not exhaustive lists and references to specific systems for individual drugs and chemicals are given in the relevant monographs in Part 2. Some of these systems use columns that are identical or very similar in terms of discriminating power (see Table .3), but are operated with different temperature programmes for specific groups of compounds. Moreover, some groups of substances are chromatographed as derivatives rather than as the parent compounds.Last modified: 4 September, 2017